Biological Safety cabinet Class 2Independent Air Flow Services can accurately test all aspects relating to the safe, efficient and standardised operation of your safety cabinets. Whether your safety cabinet operates effectively is largely contingent upon whether it meets Australian Standard specifications.

At the conclusion of our testing, you will receive a computer generated analysis report and a NATA approved Certificate of Compliance to verify that your safety cabinet meets Australian Standard specifications.

Should the testing process identify areas of underperformance, our experienced technicians will make adjustments and/or replacements, always ensuring that replacement equipment is at least equivalent in quality to the original.

To assist you with maintaining accreditation standards, our technician may also make recommendations for future action and provide a maintenance test program tailored specifically to your requirements.

Test Requirement: AS2252.2

Our testing covers:

  • AS1807.1 – Average Air Velocity & Uniformity (Clause 5.2.4)
  • AS1807.5 – Work Zone Integrity (Clause 5.2.3)
  • AS1807.6 – HEPA Filter Integrity (Clause 5.2.1)
  • AS1807.15 – Lighting Intensity (Clause 5.3.4)
  • AS1807.20 – Sound Levels (Clause 5.3.3)
  • AS1807.22 – Barrier Containment (Clause 5.2.2)
  • AS1807.23 – UV Intensity (Clause 5.3.5)
  • AS1807.26 – Determination of Air Barrier Containment – Potassium Iodide Test
  • AS ISO 14644.1:2017- Classification and air cleanliness (Particle Counts)
  • NSF/ANSI 49 Annex F.2 Downflow Velocity
  • NSF/ANSI 49 Annex F.3 Inflow Velocity (face velocity)
  • NSF/ANSI 49 Annex F.4 Air Flow Smoke Pattern Test
  • NSF/ANSI 49 Annex F.5 HEPA Filter Leak Test
  • NSF/ANSI 49 Annex F.9 Lighting Intensity
  • NSF/ANSI 49 Annex F.11 Noise level test

Independent Air Flow Services

Accreditation No: 3098

Accreditation Details: View Details

Accredited by:
The National Association of Testing Authorities